Effect of Spherical Aberration on the Optical Quality after Implantation of Two Different Aspherical Intraocular Lenses

J Ophthalmol. 2017:2017:8039719. doi: 10.1155/2017/8039719. Epub 2017 Aug 16.

Abstract

Purpose: To compare the effect of spherical aberration on optical quality in eyes with two different aspherical intraocular lenses.

Methods: 120 eyes of 60 patients underwent phacoemulsification. In patients' eyes, an aberration-free IOL (Aspira-aA; Human Optics) or an aberration-correcting aspherical IOL (Tecnis ZCB00; Abott Medical Optics) was randomly implanted. After surgery, contrast sensitivity and wavefront measurements as well as tilt and decentration measurements were performed.

Results: Contrast sensitivity was significantly higher in eyes with Aspira lens under mesopic conditions with 12 cycles per degree (CPD) and under photopic conditions with 18 CPD (p = 0.02). Wavefront measurements showed a higher total spherical aberration with a minimal pupil size of 4 mm in the Aspira group (0.05 ± 0.03) than in the Tecnis group (0.03 ± 0.02) (p = 0.001). Strehl ratio was higher in eyes with Tecnis (0.28 ± 0.17) with a minimal pupil size larger than 5 mm than that with Aspira (0.16 ± 0.14) (p = 0.04). In pupils with a minimum diameter of 4 mm spherical aberration had a significant effect on Strehl ratio, but not in pupils with a diameter less than 4 mm.

Conclusions: Optical quality was better in eyes with the aberration-correcting Tecnis IOL when pupils were large. In contrast, this could not be shown in eyes with pupils under 4 mm or larger. This trial is registered with Clinicaltrials.gov NCT03224728.

Associated data

  • ClinicalTrials.gov/NCT03224728